AERONET Instrument Troubleshooting Interface - Solutions for LOW AVERAGE CIMEL EXTERNAL BATTERY Flag
Description The instrument requires a stable 12 V supply from the external battery to operate normally. The daily mean voltage falling below a certain threshold is indicative of a failing battery. At solar-powered sites, this flag may mask the nightly lows of a failing battery due to the elevated voltage during daylight hours.
Condition(s) Average external CIMEL battery - flag if < 11.5 volts, do not include voltages < 9 volts and > 15 volts.
How To Check/Test Model 4: Voltmeter Model 5: VIEW > InsH > NEXT (disconnect solar panel before checking) Model T: See Model T Power Check
Solution Level 1 (< 5 minutes) 1. See External battery connections Model 5.jpg and External battery connections Model T.jpg 2. Check all battery and solar panel connections for corrosion (oxidation). 3. Check/reseat all connections between solar panel and control box. 4. Check/reseat connector panel. 5. Check/reseat connections going to external battery.